Wavelength |
Range |
190-1100nm |
Accuracy |
± 0.1nm |
Repeatability |
≤ 0.1nm |
|
Photometric |
Mode |
T, A, C, E |
Range |
-4.0~4.0Abs 0-200%τ |
Accuracy |
±0.2%τ(0~100%τ)
±0.002Abs(0-0.5A)
±0.004Abs(0.5-1A) |
Repeatability |
±0.001Abs(0-0.5A)
±0.002Abs(0.5-1A) |
Photometric Noise |
±0.0004A(500nm), 30min Warm-Up |
|
Operation Mode |
Manual & PC operation |
Spectrum Bandwidth |
2nm |
Baseline Flatness |
± 0.004A (200-850nm) |
Baseline Stability |
± 0.0004 Abs / h (500nm, 0 Abs), 2hr warm-up |
Optical System |
Double Beam, CT light path; Grating 1200 lines/mm holographic diffraction grating |
Light Source |
Tungsten Halogen and Deuterium lamps |
Detector |
Silicon Photodiode |
Stray Light |
≤ 0.02%T@ (measured by NaI at 220 nm) (measured at NaNO2 at 340nm) |
Display |
320X240 Graphic display LCD screen |
Software Output |
UV-Win-GLP |
Data Output |
USB 2.0 printer, USB provision for software |
Power Requirement |
Switchable 120~230VAC 50~60Hz |
Dimensions (W ×D ×H) |
670mm* 560mm*260mm |
Weight |
24Kg |