+447979485982

Atomic Absorption Spectrophotometer AAS-7100

Atomic Absorption Spectrophotometer : Atomic Absorption Spectrophotometer AAS-7100

Atomic Absorption Spectrophotometer is totally PC controlled operating system, user-friendly, ease of operation, versatile software applications and compatible for upgradation. This instrument is designed highly safety norms to operate. AAS is an advanced and affordable system with high sensitivity that generates accurate and reproducible measurements. It is an economical Flame atomic absorption spectrophotometer with high performance. It meets the requirement of limited budget

Catalog
  • Atomic Absorption Spectrophotometer is totally PC controlled operating system, user-friendly, ease of operation, versatile software applications and compatible for upgradation. This instrument is designed highly safety norms to operate. AAS is an advanced and affordable system with high sensitivity that generates accurate and reproducible measurements. It is an economical Flame atomic absorption spectrophotometer with high performance. It meets the requirement of limited budget
  • Compact, Smart looking, economic cost, upgradable
  • Totally PC-Controlled and Microcontroller based user friendly operation
  • Automated 4/6/8 lamp turret compatible to coded and non-coded HCL
  • Lamps with independent power supply for warm-up
  • Background correction with D2 lamp source
  • Manual Gas flow control
  • Ignition Method: Push Button
  • Safety Precautions: Auto shut-off Flame, Auto-shut of gas, Power Failure
  • Burner Mismatched
  • Optional: 6-Lamp Turret,8-Lamp Turret, Hydride Generator, Graphite Furnace
Wavelength
Range 190 - 900nm
Accuracy ± 0.2nm
Repeatability ± 0.1nm
Baseline Drift ≤ 0.004ABS/30min (Cu)
Spectral Bandwidth 0.1, 0.2, 0.4, 1.0 nm
Resolution ±0.2nm (Mn WL lines (279.5nm and 279.8nm)
Optical System Czerny-Turner, integrated optical platform, fully closed optical system
Grating 1800lines/ mm
Burner Head 100mm standard, Optional:50mm nitrous oxide Burner
Nebulizer High-Efficiency glass nebulizer
Baseline Drift (Flame) ≤ 0.004 ABS/30min (Cu Lamp)
Characteristic concentration Cu ≤ 0.025 µg/ml
Detection Limit Cu ≤ 0.008 µg/ml
5 PPM Cu Abs. > 8ABS
Reproducibility ≤ 0.5%
Gas Flow Control Manual
Ignition Via button on the control panel
Safety Precautions Automatically cut off gas in case of low pressure, Power Failure, Flame shut-off and Burner mismatched
Background Correction Technology D2(background signal = 1Abs, correction capability ≥ 50 times)
Dimensions 700mm*550mm*530mm
Agricultural
Geological
Education and Research
Institutes
Food
Petrochemical
Environmental
Mining
Pharmaceuticals
100mm burner+Capillary+1 lamp (Cu)+ Air Compressor + PC software
6 Lamp Turret or 8 Lamp Turret,50mm burner, Hydride Generator, Graphite Furnace System, Auto Sampler