FTIR Spectrometer–LFIS 510 |
|
- Adopting cube-corner mirrors and precision bearings with higher stability and less sensitive to vibration.
- Fully sealed damp-proof interferometer ensures higher adaptability to the environment.
- Compatible computer and various application software with IR libraries.
- The sampling compartment is wide enough for various accessories such as defused/Specular Reflection,
Horizontal ATR, Variable Angle ATR, Liquid cell and Gas cell.
|
|
|
|
| Technical Specification |
| |
| MODEL |

|
| Spectral Range |
7000 to 400 cm-1 |
| Resolution |
1 cm-1, 2 cm-1, 4 cm-1, 8 cm-1 |
| Wave Number Repeatability |
Better than ± 0.01 cm-1 |
| Scanning Speed |
0.2~2.5 cm-1/sec, Automatically optimized for detector type or manually adjustable for
specific applications |
| Beam Splitter |
KBr |
| Infrared Source |
External, air-cooled, high efficiency Reflex Sphere module |
| Detector |
DTGS, MCT (optional) |
| Data System |
Compatible computer |
| Output |
Laser printer |
| Software |
FT-IR software contains all routines needed for basic spectrometer operations |
| Catalog No. |
33260101 |
|
|
| |
|