| Model |

|
| Main Specification |
Wavelength Range |
190-900nm |
| Wavelength Accuracy |
±0.5nm |
| Resolution |
Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. |
| Baseline Stability |
0.005A/30min |
| Background Correction |
The D2 lamp background correction capability at 1A is better than 30 times.
The S-H background correction capability at 1.8A is better than 30 times. |
| Light Source System |
Lamp Number |
2 lamps (one preheating) |
| Lamp Current Adjustment |
Micro-computer control, adjustment range: 0~20mA |
| Lamp Power Supply Mode |
Powered by square pulses |
| Optical System |
Monochromator |
Single beam, Czerny-turner design grating Monochromator |
| Grating |
1800 l/mm |
| Focal Length |
277mm |
| Blazed Wavelength |
250nm |
| Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm |
Detection and Data
Processing System |
Detector |
Photomultiplier with high sensitivity and wide spectral range. |
| Electronic andMicrocomputer System |
Light energy and negative high-voltage auto-balance |
| Display Mode |
LED display of energy and measurement values, concentration direct reading |
| Read Mode |
Transient, integral, peak height, peak area. |
| Integral or sampling time is in the range of 0.1-19.9s |
| Scale Expansion |
0.1~99 |
| Data Processing Mode |
Automatic calculation of mean, standard deviation and relative standard deviation. Repeating number is in the range of 1~99 |
| Measurement Mode |
3~7 standards can be determined for curve fitting. Sensitivity correction available |
| Result Printing |
Measurement data, working curve, signal profile and analytical conditions can all be printed out |
| Instrument Self-Check |
Check current status of each function key |
| Flame Analysis |
Burner |
10cm single slot all titanium burner |
| Spray Chamber |
All plastic spray chamber |
| Nebulizer |
High efficiency glass nebulizer with metal sleeve |
| Position Adjustment |
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner |
| Gas line Protection |
Fuel gas leak alarm |
| Characteristic |
Air-C2 H2 Flame |
Cu: Characteristic concentration ≤0.025mg/L, Detection limit |
| Concentration and Detection Limit |
≤0.006mg/L |
| Function Expansion |
Flameless analysis with WF-1C Graphite Furnace System Hydride Vapor generator can be connected for hydride analysis |
| Dimensions |
1020 x 490 x 540mm |
| Weight |
80kg, Unpacked |
| Catalog No. |
05260104 |