| Model No. |

|

|

|
| Main Specification |
Wavelength Range |
190-900nm |
| Wavelength Accuracy |
±0.5nm |
| Resolution |
better than 0.3nm |
| Baseline Stability |
0.005A/30min |
| Background Correction |
The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times |
| Hollow Cathode Lamps |
Lamp Turret |
6-lamp turret automatic changeover |
4-lamp turret automatic changeover |
| Lamp Current Adjustment |
Automatic adjustment and display.Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. |
Automatic adjustment and display.Wide pulse current: 0~25mA |
| Lamp Power Supply mode |
Powered by two pulses: 400Hz wide pulse and 100Hz narrow pulse simultaneously |
400Hz square-wave pulses |
| Optical System |
Monochromator |
Single beam, Czerny-Turner design grating monochromator |
| Grating |
1800 l/mm |
| Focal Length |
277mm |
| Blazed Wavelength |
250nm |
| Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm |
| Detection and Data Processing System |
Detector |
Photomultiplier with high sensitivity and wide spectral range |
| Software |
Ms-WindowsTM operating system |
| Analytical Method |
Working Curve Method (1st order and 2nd order fitting), Standard addition method |
| Repeatable Measurement |
Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation |
| Condition Reading |
With model function |
| Result Printing |
Measurement of data and final analytical report printout |
| Flame Analysis |
Burner |
10cm single slot all titanium burner for Air-C2H2flame; 5cm single slot all titanium burner for oxygen rich flame |
10cm single slot all titanium burner for Air-C2H2flame |
| Spray Chamber |
All plastic spray chamber |
| Nebulizer |
High efficiency glass nebulizer with metal sleeve |
| Position Adjustment |
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner |
| Protection of Gas System |
Fuel gas leak protection, abnormal flame extinction alarm, air-pressure deficiency protection and explosion proof device |
| Flame Ignition |
Automatic |
| Characteristics Concentration and Detection Limit |
Normal Air C2H2 Flame |
Cu: Characteristic concentration ≤0.05mg/L, Detection limit≤0.006mg/L; |
| Oxygen Rich Air C2H2Flame |
Ba: Characteristic concentration ≤0.22mg/L |
---- |
| Hydride Generation |
Hg: detection limit ≤ 0.5μg/L |
| Dimension
(L x W x H) mm |
1020x490x540 |
| Weight |
Unpacked 80kg |
| Catalog No. |
35260104 |
35260105 |
35260106 |