Model No. |

|

|

|
| Main Specification |
Wavelength Range |
190-900nm |
| Wavelength Accuracy |
±0.25nm |
| Resolution |
Two spectral lines of Mn at 279.5nm and 279.8nm can be energy separated with the spectral bandwidth of 0.2nm and valley- peak ratio less than 30% |
| Baseline Stability |
≤0.004A/30min |
| D2 lamp Background Correction |
1A is better than 30 times |
| S-H Background Correction |
1.8A better than 30 times |
- |
| Hollow Cathode Lamps |
Lamp Turret |
6-lamp turret, Auto- alignment, fully automated scan and Peak-picking. |
4-lamp turret, Auto-alignment,
fully automated
scan and peak-picking. |
| Lamp Current Adjustment |
Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA |
| Lamp Power Supply mode |
Powered by two pulses simultaneously |
| Optical System |
Monochromator |
Single beam, Czerny turner design grating monochromator |
| Grating |
1800 l/mm |
| Focal Length |
277mm |
| Blazed Wavelength |
250nm |
| Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change |
| Atomizer |
Flame/Graphite furnace Flame/Graphite furnace atomizer auto change over |
| Detection and Data Processing System |
Detector |
Photomultiplier with high sensitivity and wide spectral range |
| Software |
MS-Windows 98/Me Operating System |
| Analytical Method |
Working Curve Method (1st order and 2nd order fitting)Sensitivity auto-correction, Standard Addition Method |
| Concentration |
Auto-calculation of concentration and content |
| Repeatable Measurement |
Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation |
| Multi-task Function |
Sequential measurement of multi-elements in one sample |
| Condition Reading |
With model function |
| Result Printing |
Measurement of Data and Final Analysis report printout, editing with Excel |
| Flame Analysis |
Burner |
10cm single slot all titanium burner |
| Spray Chamber |
All plastic spray chamber |
| Nebulizer |
High efficiency glass nebulizer with metal sleeve |
| Position Adjustment |
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner. |
| Protection of Gas System |
Fuel gas leak protection,abnormal flame extinction alarm, air-pressure deficiency protection and explosion-proof device |
| Flame Ignition |
Automatic |
| Characteristics Concentration and Detection Limit |
Normal Air C2H2 Flame |
Cu: Characteristic concentration≤0.025mg/L, Detection limit
≤0.006mg/L |
| Oxygen Rich Air C2H2Flame |
Ba: Characteristic concentration≤ 0.22mg/L |
| Al: Characteristic concentration ≤ 0.4mg/L |
--- |
| Hydride Generation |
Hg: detection limit≤0.5µg/L |
| Graphite Furnace Analysis |
Temperature Range |
20~3000ºC |
| Light-Controlled Temperature Range |
1500ºC -3000ºC, temperature auto-correction |
| Temperature Control Accuracy |
≤1% |
| Characteristic Mass |
Cd ≤0.5 ×10-12g, Cu ≤5 ×10-12g,Mo ≤1×10-11g |
| Precision |
RSD ≤4% |
| Safety Protection |
Protections for over-heating,over-current, and less cooling water and argon pressure monitoring functions |
Dimension
(L x W x H) mm |
Main Unit |
1020 (L) × 492 (W) × 537 (H)mm |
| Graphite Furnace |
420 (L) × 420 (W) × 460 (H) mm |
| Weight |
Main Unit |
80kg , Unpacked |
| Graphite Furnace |
50kg |
| Catalog No. |
35260101 |
35260102 |
35260103 |