Spectrophotometers Live Help
Email ID Password

Atomic Absorption Spectrophotometer AAS-010

Atomic Absorption Spectrophotometer Model AAS-010 is widely used in Pharma, Biotech & Medical research applications. It is fully automated and provided with an automatic change over of spectral bandwidth with a choice of 5 slit sizes. Its unique design of the graphite furnace reduces the chemical interference , memory effects and leads to uniform heating of the graphite electrode.

Features:

  • Fully Automated and Innovative design, with the flame and graphite furnace integrated into one instrument.
  • Motorized 8 hollow cathode lamp turret with two methods of background correction.
  • Spectral bandwidth is automated & available with a choice of 5 slit sizes.
  • The unique design of the graphite furnace reduces the chemical interference and memory effects and leads to uniform heating of the graphite electrode. The computer controlled heating program allows the user to select the best heating program for the analysis. The optical temperature during the atomization stage ensures the rapid heating and rapid analysis. This helps to extend the life of the graphite tube and enhances analytical accuracy.
  • The flame conditions are continuously monitored and should the flow rates change an audible alarm sounds. The pressure of the support gas (oxidant) is monitored constantly. If the pressure changes then the flow of the fuel gas will be stopped and the flame will be safely extinguished. A sensor monitors the level of liquid in the drain and will prevent ignition if too low. The flame will also be extinguished if the level of liquid in the drain, changes significantly, the argon pressure for the graphite furnace is constantly monitored and should it change, the heating cycle for the graphite electrode will immediately cease and the graphite electrode will be de energized. Cooling water flow rates for the graphite furnace are also Monitored for changes and should Changes occur the heating program will cease. If the graphite tube should fracture during the heating program, the heating will ceasethe ignition of the flame is computerized controlled and the various safety Interlocks offer a very safe Operating system.

Specifications:

Main Specification:

Wavelength Range:
190nm
900nm
Wavelength Accuracy:
± 0.25 nm
Wavelength Repeatabilty:
0.15 nm
Resolution:
2μg/ml Cu gives approximately 0.38 Abs
Baseline Stability 0.005A/ 30 min:
0.005A/ 30 min
Background correction:
Deuterium lamp1.0Abs
Self Reversal:
1.0 Abs

Optical System:

Monochromator:
Czerny - Turner Configuration Single Beam
Spectral Bandwidth:
0.1nm, 0.2 nm,0.4 nm,1.0 nm,2.0 nm 5 steps with automatic changeover

Flame Atomizer:

Sensitivity:
(Cu) 0.03 ug /ml/1%
Burner Head:
Titanium alloy burner
Nebulizer:
High efficiency glass nebulizer
Atomization Chamber:
Corrosion resistant material
Position adjustment:
Automatic changeover of Flame and Furnace, Automatic setting of optimum height for Flame burner Automatic ignition and of mixing air - acetylene gas with safety control

Detection and Data Processing System:

Analytical method:
Flame, Graphite furnace and Hydride
Determination Methods:
Calibration curves using 1st, 2nd and 3rd order of fit, standard addition method
Repetitions:
1-20 with calculations of average, SD and RSD

Light Source System:

Light Source:
8 hollow cathode lamp turrets with 2 simultaneously lit , one lamp pre heated

Graphite Furnace:

Character Value:
( Cd) 0.5 pg
Temperature range:
Ambient - 2650 °C
Heating:
Voltage feedback control when drying and ashing, optical temperature control when atomizing
Heating Program:
Upto 9 steps with choices of ramp, temperature increase and full power heating

Dimensions:

Main unit:
110 x 50 x 45
Graphite furnace:
50 cm x 50 cm x 45 cm

Product Details:

Add to Cart
Product Name:
Atomic Absorption
Model:
Atomic Absorption Spectrophotometer AAS-010
Unit Price:
Send request for Price: info@SpectroPhotoMeters.com
Quantity:
 
Add to Cart


Other Models:

   Atomic Absorption Spectrophotometer AAS-001

   Atomic Absorption Spectrophotometer AAS-003

   Atomic Absorption Spectrophotometer AAS-004

   Atomic Absorption Spectrophotometer AAS-007

   Atomic Absorption Spectrophotometer AAS-008

   Atomic Absorption Spectrophotometer AAS-009

Atomic Absorption Spectrophotometer AAS-010   
Atomic Absorption Spectroscopy   
Atomic Absorption   
Flame Atomic Absorption   
Graphite Furnace Analysis   
Flame Spectrophotometer   
Graphite Furnace Spectrophotometer   
Spectrophotometer.   

Atomic Absorption Spectrophotometer Model AAS-009 features high-sensitivity analysis, flexible system configurations, a number of safety features, and a compact footprint, ensuring efficient and comfortable operation in your laboratory.Low detection limits and simple, rapid testing methods make these sensitive atomic absorption spectrophotometer ideal for environmental testing, toxicology, and quality control labs.

Sign up for Product Updates

 
Send a Request
Name:
Mail ID:
Subject:
Comments: